ISCEV and IPS guideline for the full-field stimulus test (FST)
Date
2024-01-18Author
Jolly, JK
Grigg, JR
McKendrick, AM
Fujinami, K
Cideciyan, AV
Thompson, DA
Matsumoto, C
Asaoka, R
Johnson, C
Dul, MW
Artes, PH
Robson, AG
Subject
Threshold Sensitivity Scotopic Photopic Clinical standards International Society for Clinical Electrophysiology of Vision Imaging and Perimetry Society
Metadata
Show full item recordAbstract
The full-field stimulus test (FST) is a psychophysical technique designed for the measurement of visual function in low vision. The method involves the use of a ganzfeld stimulator, as used in routine full-field electroretinography, to deliver full-field flashes of light. This guideline was developed jointly by the International Society for Clinical Electrophysiology of Vision (ISCEV) and Imaging and Perimetry Society (IPS) in order to provide technical information, promote consistency of testing and reporting, and encourage convergence of methods for FST. It is intended to aid practitioners and guide the formulation of FST protocols, with a view to future standardisation.
Collections
Publisher
Springer Science and Business Media LLC
Place of Publication
Netherlands
Journal
Documenta Ophthalmologica
Volume
148
Issue
1
Pagination
3-14